140 Chapter 11
Bolus stopped
If this alarm should occur, it is
very important to check bolus
history to review how much of
the bolus was actually delivered.
Reprogram a bolus with the
amount not delivered, if
required.
Failed batt test
The pump is designed to test
each new battery installed in
the pump. This is a safety fea-
ture to make sure that a battery
that already has low voltage is
not used. If the battery does not
pass the test this alarm will
occur.
Always make sure that you
install a NEW battery into the pump.
This alarm will also happen if a used battery is installed in the
pump or if the installed battery is removed and then put back in.
MiniMed
BOLUS
ESC,ACT to clear
Loose battery cap?
8.35A
STOPPED
Pump dropped or
bumped?
Check bolus history,
Reprogram bolus
if required
BATT TEST
ESC,ACT to clear
Replace with
8.35A
FAILED
Delivery stopped
new battery