How to Perform a Gas Test
1. Ensure the G-TECTA™ 4GP instrument is switched
on and in normal operating mode.
2. Fit the flow cap onto the front of the sensors and
attach the hose from the regulator. Attach the
outlet hose to ‘vent gas away’ – do not extend
this hose, restrict or allow kinks.
3. Swipe the Gas Test Key past the display, adjacent
to the LED lens. The G-TECTA™ 4GP will activate
the Gas Test Mode and show “TEST” and “x” next
to the sensors on the display.
4. G-TECTA™ 4GP will display a progress bar.
5. Apply the gas whilst the progress bar
is counting down.
6. Wait for the progress bar to complete
(40 seconds). G-TECTA™ 4GP will display a
“3” or “x” against each sensor and a master
“3” or “x” to indicate PASS or FAIL
In the event G-TECTA™ 4GP displays FAIL, please
see the troubleshooting guide (Page 17) or
contact The Linde Group.
7. To abort the gas test press the mode button at
any time whilst the test is in progress.
Instrument Calibration
Calibration of the gas sensor should be performed
regularly every 6 months. Over time the accuracy of
the gas sensor can drift outside of what is considered
a safe range and regular calibration is essential to
ensure accuracy of gas readings.
The accuracy of G-TECTA™ 4GP
instruments should be checked periodically with
known-concentration of calibration gas. Failure to
check accuracy can lead to inaccurate and potentially
dangerous readings
There are three methods of instrument calibration
available. G-TECTA™ 4GP calibration can be
performed directly on each instrument (one button
calibration), by utilising G-TECTA™ PC software or by
using the Linde G-DOCKA™ Hub. Certified calibration
gases should be used.
How to Perform a One Button Calibration Test
1. Ensure you are in clean air and the instrument
is in normal operating mode. Double click the
mode button and select “Zero” from the options
menu by double clicking on the “Zero” icon.
G-TECTA™ 4GP will perform an Auto Zero.
To perform a one button calibration,
complete the next steps within 15 minutes
of completing the Auto Zero.
2. Fit the flow cap onto the front of the sensors and
attach the hose from the regulator. Attach the
outlet hose to ‘vent gas away’ – do not extend
this hose, restrict or allow kinks.
3. Swipe the Gas Test Key past the display, adjacent
to the LED lens. The G-TECTA™ 4GP will activate
the Calibration Mode and show “CAL” and “????”
on the display.
4. Press the mode button within 10 seconds to
confirm calibration.
Note: If the mode button is not pressed within
10 seconds then the instrument mode will change to
Gas Test Mode.
5. G-TECTA™ 4GP will display a progress bar and an
“x” will appear against each gas type.
6. Apply the gas whilst the progress bar is
counting down.
Note: You can abort the calibration test by
pressing the mode button at any time whilst
the test is in progress.
7. The G-TECTA™ 4GP instrument will adjust the
value on the gas sensor to match the stored
calibration gas value within the sensor module.
8. G-TECTA™ 4GP will display a “3” or “x” against
each sensor to indicate PASS or FAIL.
9. If any of the G-TECTA™ 4GP sensors do not
calibrate successfully, the gas type will remain
marked with a “x” and the warning messages
“Gas Test Failed” and “Send for Calibration” will
be displayed. The G-TECTA™ 4GP must be sent
to a Linde Group Authorised Service centre for
re-calibrating or sensor replacement.
Note: Gas Test pass and fail, and calibration
pass, fail and values are stored in the Event log.
Note: Some sensors may not be calibration enabled.
Oxygen sensors are by default not calibration
enabled as they are calibrated in fresh air by zeroing.
One button calibration will
check for small drifts of stored calibration value.
The Linde Group recommends G-TECTA™ 4GP is sent
for full certified calibration at six monthly intervals.
G-TECTA 4GP Instrument with
Ozone (O
3
) Sensor
Due to the reactive nature of ozone (O
3
) special
procedures should be followed when calibrating gas
detectors that incorporate an ozone sensor.
Fittings
For ozone calibrations all fittings and pipe work
should be stainless steel, brass, aluminium or PTFE.
Fittings or pipe work in other plastic materials such
as Tygon must not be used.
Gas flow rate
It is important that the flow rate is set correctly as
otherwise pressure effects may distort calibration
values and prevent the gas detector from working
correctly. To calibrate, the flow rate should be set
between 0.8-1.0 litres/minute (0.03-0.04 cubic
feet/minute)
Normal safety precautions for handling ozone
should always be observed along with any special
instructions that accompany the calibration gas
cylinder or generator being used.
For calibrating the ozone sensor, the special flow
cap assembly with four inlets (article number
AS4GP3021) must be used. The Ozone sensor must
always be at position 3, as marked on the flow cap.
Note: The standard flow cap assembly supplied with
the instrument must be used for calibrating the other
gas sensors
* G-TECTA™ PC Software can be downloaded free of charge from www.g‑tecta.com
Commentaires sur ces manuels