Dilution Manual
Cannonball 3
Multi Gas Detector
Sperian Instrumentation
651 South Main Street
Middletown, CT 06457
Tel (860) 344-1079 (800) 711-6776
Fax (860) 344-1068
Version 1.11 6DEC2010 p/n 13-186
Vue de la page 0
1 2 3 4 5 6 ... 20 21

Résumé du contenu

Page 1

Dilution Manual Cannonball 3 Multi Gas Detector Sperian Instrumentation 651 South Main Street Middletown, CT 06457 Tel (860) 344-1079 (800) 7

Page 2

10sensor is replaced in a dilution-capable Cannonball3. 3.2 Automatic calibration in dilution mode Automatic calibration in dilution mode may only b

Page 3 - Table of Contents

11 Dilution Fresh Air Cal Please Wait….. 8 The Cannonball 3 will then proceed to the automatic span calibration. SPAN CALIBRATION MODE

Page 4 - Warnings

12 LEL Calibrated to 50 CO Calibrated to 50 H2S Calibrated to 25 The Cannonball 3 will then display the following screen, which shows the fin

Page 5

13balloon may lead to inaccurate and potentially dangerous gas readings. Use of non-standard calibration gas and/or calibration kit components when

Page 6 - Monitoring

143.3 Manual dilution calibration procedures Cannonball 3 instruments with dilution capability have five different manual calibration options in the

Page 7 - 20.9 0

15 CALIBRATION MENU GAS VALUES SPAN CAL HISTORY O2 ZERO CAL FRESH AIR REMINDER EXIT 3. Use the navigation arrows to highlight O2 ZERO CAL and

Page 8

16DILUTION PUMP ON O2 Sensor Reading 0.0 Press MODE to calibrate 7. The oxygen reading will drop immediately and should approach

Page 9 - 3. Calibration in dilution

17 Make sure the calibration balloon is filled with the appropriate calibration gas before attempting to calibrate the Cannonball 3. Use of imprope

Page 10

18 Dilution Fresh Air Cal Please Wait….. 14  Fresh Air Calibration Completed Upon completion of the manual dilution fresh air

Page 11

19raise or lower the readings to match the concentration printed on the calibration cylinder label. DILUTION SPAN CALIBRATION LEL 0 CO 0.0 H2S

Page 12

2 THE CANNONBALL 3 PERSONAL PORTABLE GAS DETECTOR HAS BEEN DESIGNED FOR THE DETECTION OF OXYGEN DEFICIENCIES, FLAMMABLE GAS, AND TOXIC VAPOR ACCUMU

Page 13

20Appendix A: Calibration FrequencyOne of the most common questions that we are asked at Sperian is: “How often should I calibrate my gas detector?”

Page 14

21Sperian Standard Warranty Gas Detection Products General Sperian Protection Instrumentation, LLC (hereafter Sperian) warrants gas detectors, sensor

Page 15

3Table of Contents Signal Words ...4 Warnings...

Page 16

4Signal Words The following signal words, as defined by ANSI Z535.4-1998, are used in the Cannonball 3 Dilution and Reference Manuals. indicates an

Page 17

5Sperian Instrumentation offers calibration kits and long-lasting cylinders of test gas specifically developed for easy Cannonball 3 calibration. Cu

Page 18

61. Overview of Dilution Monitoring The Cannonball 3 sample dilution system allows the user to accurately sample combustible gases from otherwise in

Page 19

7alarm set points are automatically changed. The user will have to acknowledge the new oxygen alarm set points by pressing the MODE button. Instrume

Page 20

8 Checking for Dilution Pump…..  Setting O2 Alarms For Dilution Mode LOW CEIL O2 OFF 5.0 MODE=Acknowledge 4. Pre

Page 21 - Sensor Warranty Periods

9 ALARMS MENU CUSTOM LATCH DEFAULT TEMPERATURE EXIT 2. Use the navigation arrows to highlight CUSTOM and press the MODE button. The curre